Agilent Technologies 35670-90066 manual • Press the following keys to start the test

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 67
Image 67
• Press the following keys to start the test:

Agilent 35670A

Verifying Specifications

 

To run the program in semiautomated mode

Press the following keys and type in the printer address when the program prompts you:

 

[ TEST CONFIG ]

 

[ PRINTER ADDRESS ]

 

[ PROCEDURE ]

 

[ OP_VERIFY ] or [ PERFORMAN ]

 

[ STOP AFTER ]

 

[ LIMIT FAILURE ] or [ NONE ]

 

[ RETURN ]

 

Press the following keys to start the test:

 

[ START TESTING ]

 

[ START BEGINNING ]

 

When you select [ START BEGINNING ], the data is written to a file on the disk and

 

printed only after all tests are done. When you select [ START MIDDLE ] or

 

[ ONE TEST ], the data is printed immediately after each measurement.

 

Follow the directions on the display.

 

 

Warning

During the test, the program prompts you to change the test equipment

 

connections. Always turn the ac calibrator output to OFF or STANDBY before

 

changing test equipment connections. The ac calibrator can produce output

 

voltages that could result in injury to personnel.

 

The directions on the display briefly tell you how to connect test equipment. For

 

 

detailed illustrations of equipment setup, see the setup illustrations starting on page

 

3-13.

 

If you want to pause the program and return the Agilent 35670A Dynamic Signal

 

Analyzer to front panel control, press [ BASIC ]. To continue the program, press

 

[ BASIC ] [ DISPLAY SETUP ] [ LOWER ] [ RETURN ] [ CONTINUE ]. If you changed

 

any instrument setup states, press [ RESTART TEST ] instead of [ CONTINUE ]to

 

ensure accurate measurement results.

3-9

Page 67
Image 67
Agilent Technologies 35670-90066 manual • Press the following keys to start the test