Troubleshooting the Analyzer

 

To troubleshoot input failures on four channel analyzers

Step 2. Exchange the Input assemblies.

GSet the power switch to off ( O ).

Exchange the Input assembly in the lower slot with the Input assembly in the upper slot.

Reconnect the cables to the A5 Analog assembly.

Set the power switch to on ( l ).

Press the following keys:

[ System Utility ]

[ CALIBRATN ]

[ AUTO CAL OFF ] [ Input ]

[ ALL CHANNELS ]

[ CH* FIXED RANGE ] 1

[ Vpk ]

[ System Utility ]

[ MORE ]

[ SELF TEST ] [ TEST LOG ]

[ Rtn ]

[ FUNCTIONL TESTS ] [ INPUTS ]

[ ALL ]

[ CONTINUE ] [ Rtn ]

[ Rtn ]

[ QUICK CONF TEST ]

If the same channel fails as failed before the exchange, the A5 Analog assembly is probably faulty.

If a different channel now fails, the A2 Input assembly for the failing channel is probably faulty.

The Input assembly for channel 1 and 3 is in the lower slot. The Input assembly for channel 2 and 4 is in the upper slot.