Agilent 35670A

 

Verifying Specifications

 

 

Measurement Uncertainty

Performance Test

Using Recommended Test

Using Other Test Equipment

 

Equipment

 

 

 

 

 

 

Measurement Uncertainty Ratio

Measurement Uncertainty Ratio

Amplitude Linearity

13dBVrms

−1 dBVrms −15 dBVrms −29 dBVrms −43 dBVrms −53 dBVrms

±0.0020 dB

>10:1

±0.0020 dB

>10:1

±0.0026 dB

>10:1

 

±0.0046 dB

>10:1

>10:1

±0.0096 dB

>10:1

±0.0255 dB

 

A-Weight Filter

 

 

10 Hz

0.016 dB

>10:1

31.62 Hz

0.012 dB

>10:1

100 Hz

0.012 dB

>10:1

1 kHz

0.011 dB

>10:1

10 kHz

0.011 dB

>10:1

25.120 kHz

0.012 dB

>10:1

 

 

 

 

Channel Match

 

 

magnitude

±0.00001 dB

>10:1

phase

±0.01 mdeg

>10:1

 

 

 

 

Frequency Accuracy

±6.25 ppm

4.8:1

 

 

 

Anti-Alias Filter

 

 

<100 kHz

± 0.1 dB

>10:1

<1 MHz

± 0.3 dB

>10:1

 

 

 

 

Input Coupling

± 0.001 dB

>10:1

 

 

 

Harmonic Distortion

 

 

using HP 339A

±0.184 dB

4.46:1

using HP 3326A with filter

±0.92 dB

10:1

 

 

 

 

Intermodulation Distortion

±0.83 dB

10:1

 

 

 

Cross Talk

 

 

channel to channel

± 0.1 dB

>10:1

source to input

±1.34 dB

6:1

 

 

 

 

Single Channel Phase Accuracy

± 0.25 deg †

>10:1

 

 

 

External Trigger

280 mVpk

3.6:1 ‡

 

 

 

Tach Function

330 mV

6:1

 

 

 

Input Resistance

±17Ω

>10:1

 

 

 

ICP Supply

17

 

oven circuit voltage

±320 mV

>10:1

current

±132<M > mA

>10:1

 

 

 

 

3-57