Agilent 35670A

Voltages and Signals

 

 

A99 Motherboard

FSELAn

Fast Bus Asynchronous Select — This line is low when an asynchronous fast bus cycle is in

 

operation.

 

FSELSn

Fast Bus Synchronous Select — This line is low when a synchronous fast bus cycle is in

 

operation. The A6 Digital assembly uses this signal to enable I/O to its gate arrays and to the

 

A5 Analog assembly’s source attenuator DAC and dc offset DAC.

 

H10MHZ

10 MHz Clock — This is a 50% duty cycle, 9.961472 MHz clock. This clock is H20MHZ

 

divided by 2.

 

H20MHZ

20 MHz Clock — This is a 50% duty cycle, 19.922944 MHz clock. This clock provides the

 

timing for the analyzer.

 

HRNGA

Half Range A — In both the two channel and four channel analyzer, a high on this line turns on

 

the channel 1 half range LED. This line goes high when the A1 or A2 Input assembly detects

 

that the amplitude of the channel 1 input signal reached half the set range.

 

HRNGB

Half Range B — In a two channel analyzer, a high on this line turns on the channel 2 half range

 

LED. In a four channel analyzer, a high on this line turns on the channel 3 half range LED.

 

This line goes high when the A1 Input assembly detects that the amplitude of the channel 2

 

input signal reached half the set range or when the A2 Input assembly detects that the

 

amplitude of the channel 3 input signal reached half the set range.

 

HRNGC

Half Range C — In a four channel analyzer, a high on this line turns on the channel 2 half

 

range LED. This line goes high when the A2 Input assembly detects that the amplitude of the

 

channel 2 input signal reached half the set range. This line is only used in four channel

 

analyzers.

 

HRNGD

Half Range D — In a four channel analyzer, a high on this line turns on the channel 4 half

 

range LED. This line goes high when the A2 Input assembly detects that the amplitude of the

 

channel 4 input signal reached half the set range. This line is only used in four channel

 

analyzers.

 

HSYNC

Horizontal Synchronization — A high on this line causes the external monitor to do a

 

horizontal retrace. The Motherboard buffers this signal and routes it to the EXT MONITOR

 

connector.

 

IFCn

Interface Clear — This line is only used during the analyzer’s development.

 

IMODE

Input Mode — This line indicates the A98 Power Supply assembly’s input power mode. When

 

this line is high, the Power Supply assembly is operating on ac power. When this line is low,

 

the Power Supply assembly is operation on dc power.

 

INTHPIBn

GPIB Controller Interrupt — A low on this line interrupts the A7 CPU assembly. This line is

 

controlled by the A10 Rear Panel assembly’s GPIB controller.

 

INTONIXn

Parallel Port Interrupt — A low on this line interrupts the A7 CPU assembly. This line is

 

controlled by the A10 Rear Panel assembly’s parallel port controller.

 

KEYVALID

Key Valid — A high on this line interrupts the A7 CPU assembly. This line goes high when a

 

key is pressed on an external keyboard.

 

9-31