Troubleshooting the Analyzer

Agilent 35670A

To perform self tests

 

Step 2. Compare the analyzer’s self-test results to the following table.

When the tests have finished, press the following keys: [ Rtn ]

[ TEST LOG ]

Press the [ PREVIOUS PAGE ] softkey until the first page of test log is

G

displayed.

To print the test log to a GPIB printer, press the following keys:

[ Local/GPIB ]

[ SYSTEM CONTROLLR ]

[ PRINTER ADDRESS ]

(printer address)

 

[ ENTER ]

 

[ System Utility ]

 

[ MORE ]

 

[ SELF TEST ]

 

[ TEST LOG ]

 

[ Plot/Print ]

 

[ <F"Times""P10 >PLOT/PRNT DEVICE ]

 

(device type)

 

[ Rtn ]

 

[ PLOT/PRNT DESTINATN ]

 

[ OUTPUT TO GPIB ]

 

[ Rtn ]

 

[ START PLOT/PRNT ]

 

If the analyzer did not complete the tests (analyzer locks up), go to page 4-37

 

‘’To troubleshoot self-test lockup failures.’’

 

If the failure is intermittent and the analyzer passed all self tests, go to page

 

4-40‘’To troubleshoot intermittent failures.’’

 

If the analyzer completed the tests, compare the analyzer’s test log to the

 

following table.

 

If the analyzer’s test log matches more than one entry on the table, use the entry

 

closest to the beginning of the table. The table lists the probable faulty assembly or

 

assemblies and any recommended adjustment or troubleshooting procedure to do

 

before replacing the assembly. If both an adjustment and a test are recommended,

 

do the adjustment first.

 

For additional information on the self tests, see ‘’Self-Test Descriptions’’ on page

 

10-10.

4-32