Time Domain

Agilent 35670A

Specifications

 

Time Domain

Time Domain

Specifications apply in histogram/time mode, unfiltered time display

DC amplitude accuracy

±5.0 % fs

 

 

Rise time of −1 V to 0 V test pulse

<11.4 ms

 

 

Settling time of −1 V to 0 V test pulse

<16 ms to 1%

 

 

Pulse aberrations (peak overshoot)

<3 %

of −1 V to 0 V test pulse

 

Peak aberration relative to the mode-to-mode

 

difference (most common values)

 

 

 

Sampling period

 

 

 

1 channel mode

3.815 ms (1/262144 Hz) to 2 s in 2⋅ steps

2 channel mode

7.629 ms (1/131072 Hz) to 4 s in 2⋅ steps

4 channel mode (option AY6 only)

15.26 ms (1/65536 Hz) to 8 s in 2⋅ steps

Trigger

Trigger modes

Internal trigger

 

External trigger

 

Source trigger

 

GPIB trigger

 

 

Maximum trigger delay

 

 

 

Post trigger

8191 seconds

Pre trigger

8191 sample periods

No two channels can be further than ±7168 samples

 

from each other.

 

 

 

External trigger maximum input

±42 Vpk

 

 

External trigger range

 

 

 

Low range

−2 V to +2 V

High range

−10 V to +10 V

 

 

External trigger resolution

 

 

 

Low range

15.7 mV

High range

78 mV

1-9