[ Disp Format ]

 

Internal Test Descriptions

Agilent 35670A

 

Calibration Routine Description

 

 

[ Inst Mode ]

 

 

[ 2 CHANNEL ]

 

 

[ Input ]

 

 

[ CHANNEL 1 RANGE ]

 

 

1

 

 

[ dBVrms ]

 

 

[ CHANNEL 2 RANGE ]

 

 

1

 

 

[ dBVrms ]

 

 

[ Disp Format ]

 

 

[ UPPER/LOWER ]

 

 

[ System Utility ]

 

 

[ CALBRATIN ]

 

 

[ SAVE CH1 CAL TRACE ]

 

 

[ INTO D1 ]

 

 

[ SAVE CH2 CAL TRACE ]

 

 

[ INTO D2 ]

 

 

[ Meas Data ]

 

 

[ MORE ]

 

 

[ DATA REGISTER ]

 

 

[ D2 ]

 

 

[ Active Trace ]

 

 

[ D1 ]

 

 

[ Scale ]

 

 

[ Y PER DIV (DECADES) ]

 

 

1

 

 

[ ENTER ]

 

 

[ CENTER REFERENCE ]

 

 

0

 

 

[ ENTER ]

 

 

[ Active Trace ]

 

 

[ Y PER DIV (DECADES) ]

 

 

1

 

 

[ ENTER ]

 

 

[ CENTER REFERENCE ]

 

 

0

 

 

[ ENTER ]

 

 

 

Note

Display A shows the message D1 CAL CHAN 1 and display B shows the message

 

D2 CAL CHAN 2.

 

The calibration correction curves should be contained within ±5 dB. The ±5 dB limit applies to all ranges at full span.

10-8