Specifications

Specifications

Agilent 35670A

Single Channel Amplitude

 

Single Channel Amplitude

Absolute amplitude accuracy (FFT)

±2.92% (0.25 dB) of reading

(A combination of full scale accuracy, full

±0.025% of full scale

scale flatness, and amplitude linearity.)

 

 

 

FFT full scale accuracy at 1 kHz (0 dBfs)

±0.15 dB (1.74%)

 

 

FFT full scale flatness (0 dBfs) relative to 1

±0.2 dB (2.33%)

kHz

 

 

 

FFT amplitude linearity at 1 kHz

±0.58% (0.05 dB) of reading

Measured on +27 dBVrms range with time

±0.025% of full scale

average, 0 to −80 dBfs.

 

 

 

Amplitude resolution (16 bits less 2 dB over-range)

0.0019% of full scale (typical)

with averaging

 

 

 

Residual dc response

<−30 dBfs or -66dBVdc (0.5 mVdc) (whichever is

FFT mode frequency display

greater)

(excludes A-weight filter)

 