Agilent 35670A

Verifying Specifications

 

To set up the input resistance test

To set up the input resistance test

Performance Test only

This test verifies that the Agilent 35670A meets its input resistance specification. In this test, a digital multimeter directly measures the input resistance of each channel. The digital multimeter is set to the 1 Mrange.

1 2 ch

2 2 ch

1 4 ch

3-39

Page 97
Image 97
Agilent Technologies 35670-90066 manual To set up the input resistance test