Agilent Technologies 35670-90066 Amplitude Accuracy continued, Verifying Specifications, 8 of

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 168
Image 168
Manual background

 

Verifying Specifications

 

 

Agilent 35670A

 

Operation Verification Test Record - Four Channel

 

 

 

Serial Number:_______________________Report Number:____________________

 

Test Date:___/___/___

 

 

 

 

 

Amplitude Accuracy (continued)

 

 

 

 

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

 

(dBfs)

(dBfs)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

35 dBVrms, Ch 2

35.15

34.85

 

 

 

 

 

 

 

 

 

35 dBVrms, Ch 3

35.15

34.85

 

 

 

 

 

 

 

 

 

35 dBVrms, Ch 4

35.15

34.85

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 1

27.15

26.85

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 2

27.15

26.85

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 3

27.15

26.85

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 4

27.15

26.85

 

 

 

 

 

 

 

 

 

11 dBVrms, Ch 1

11.15

10.85

 

 

 

 

 

 

 

 

 

11 dBVrms, Ch 2

11.15

10.85

 

 

 

 

 

 

 

 

 

11 dBVrms, Ch 3

11.15

10.85

 

 

 

 

 

 

 

 

 

11 dBVrms, Ch 4

11.15

10.85

 

 

 

 

 

 

 

 

 

1 dBVrms, Ch 1

0.85

1.15

 

 

 

 

 

 

 

 

 

1 dBVrms, Ch 2

0.85

1.15

 

 

 

 

 

 

 

 

 

1 dBVrms, Ch 3

0.85

1.15

 

 

 

 

 

 

 

 

 

1 dBVrms, Ch 4

0.85

1.15

 

 

 

 

 

 

 

 

 

9 dBVrms, Ch 1

8.85

9.15

 

 

 

 

 

 

 

 

 

9 dBVrms, Ch 2

8.85

9.15

 

 

 

 

 

 

 

 

 

9 dBVrms, Ch 3

8.85

9.15

 

 

 

 

 

 

 

 

 

9 dBVrms, Ch 4

8.85

9.15

 

 

 

 

 

 

 

 

 

19 dBVrms, Ch 1

18.85

19.15

 

 

 

 

 

 

 

 

 

19 dBVrms, Ch 2

18.85

19.15

 

 

 

 

 

 

 

 

 

19 dBVrms, Ch 3

18.85

19.15

 

 

 

 

 

 

 

 

 

19 dBVrms, Ch 4

18.85

19.15

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 1

26.85

27.15

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 2

26.85

27.15

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 3

26.85

27.15

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 4

26.85

27.15

 

 