Troubleshooting the Analyzer

Agilent 35670A

To troubleshoot performance test failures

 

To troubleshoot performance test failures

With the exception of the Quick Confidence test, all functional self tests must pass before the following table is valid.

Step 1. If the analyzer failed a performance test, compare the failing performance test

G

to the following table.

 

 

 

 

 

 

If more than one performance test is failing, use the entry that is closest to the

 

beginning of the table. The table lists the probable faulty assembly or assemblies and

 

any recommended adjustment or troubleshooting procedure to do before replacing the

 

assembly. If both an adjustment and a test are recommended, do the adjustment first.

 

Failing Performance Test Troubleshooting Guide

 

Failing Performance Test

Probable Faulty Assembly

Adjustment

Troubleshooting Test

 

 

(in order of probabilty)

 

 

 

 

 

 

 

 

 

 

 

 

DC offset

 

A1/A2 Input

Input dc offset, page 5-10

 

one channel or

 

 

ADC gain, offset, and

 

channel 1 and 3 or

 

 

reference, page 5-7

 

channel 2 and 4

 

 

 

 

 

 

 

 

 

DC offset

 

A5 Analog

Input dc offset, page 5-10

 

all channels

 

 

ADC gain, offset, and

 

 

 

 

reference, page 5-7

 

 

 

 

 

 

Amplitude accuracy

 

A1/A2 Input

Input dc offset, page 5-10

Input and ADC, page 4-51

one channel

 

A5 Analog

ADC gain, offset, and

Four channel, page 4-54

 

 

 

reference, page 5-7

 

 

 

 

 

 

Amplitude accuracy

 

A2 Input

Input dc offset, page 5-10

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

 

Amplitude accuracy

 

A5 Analog

Input dc offset, page 5-10

 

all channels

 

 

ADC gain, offset, and

 

 

 

 

reference, page 5-7

 

 

 

 

 

 

Flatness

 

A1/A2 Input

Input dc offset, page 5-10

Input and ADC, page 4-51

one channel

 

A5 Analog

ADC gain, offset, and

Four channel, page 4-54

 

 

 

reference, page 5-7

 

 

 

 

 

 

Flatness

 

A2 Input

Input dc offset, page 5-10

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

 

Flatness

 

A5 Analog

Input dc offset, page 5-10

 

all channels

 

 

ADC gain, offset, and

 

 

 

 

reference, page 5-7

 

4-42

Page 218
Image 218
Agilent Technologies 35670-90066 manual To troubleshoot performance test failures