Troubleshooting the Analyzer

Agilent 35670A

To perform self tests

 

Step 3. Determine the probable faulty assembly and next test by comparing the analyzer’s symptoms to the following table.

G

Failure

Probable Faulty

Assembly

Next Test

 

 

 

 

Disk drive

A100 Disk Drive

 

Disk drive, page 4-57

 

Flexible disk

 

 

 

Disk drive cable

 

 

 

A7 CPU

 

 

 

 

 

 

GPIB

A10 Rear Panel †

 

 

Serial port

A10

Rear Panel †

 

 

 

 

 

 

 

 

 

 

 

 

Parallel

A10

Rear Panel †

 

 

 

 

 

 

 

 

 

External trigger

A10

Rear Panel

 

Trigger failures, page 4-62

 

 

A5 Analog

 

 

 

 

 

 

 

 

 

Tachometer

A10

Rear Panel

 

Tachometer, page 4-24

 

 

A6

Digital

 

 

 

 

 

 

 

 

 

Source

A5

Analog

 

Source and calibrator, page 4-45

 

 

A6

Digital

 

 

 

 

 

 

 

 

Autorange

A1/A2 Input

 

Auto-range, page 4-59

 

Overrange

A5

Analog

 

 

 

 

 

 

 

 

 

External keyboard

A10

F200 fuse

 

DIN connector, page 4-61

 

 

External Keyboard

 

 

 

 

A10

Rear Panel

 

 

 

 

 

 

 

 

 

External monitor

A7

CPU

 

 

 

 

 

 

 

 

 

Microphone power

A77

Microphone

 

Microphone power, page 4-69

 

Microphone adapter

A5

Analog

 

 

 

 

 

 

 

 

 

 

Performance test

 

 

 

 

Performance test, page 4-42

 

 

 

 

 

 

 

 

Intermittent failure

 

 

 

 

Intermittent, page 4-40

The circuits for the output ports are located on the A10 Rear Panel assembly except for a few output buffers on the A7 CPU assembly. If replacing the A10 Rear Panel assembly does not fix the failure, the A7 CPU assembly is probably faulty.

For additional information on the self tests, see ‘’Self-Test Descriptions’’ starting on page 10-10.

4-36