Equipment Required

Agilent 35670A

Troubleshooting the Analyzer

Equipment Required

See ‘’Recommended Test Equipment’’ starting on page 1-17for tables listing recommended equipment. Any equipment which meets the critical specifications given in the tables may be substituted for the recommended model.

Troubleshooting Hints

Incorrect bias supply voltages can cause false diagnostic messages. In most troubleshooting procedures, the power supply voltages are not checked. If you suspect incorrect supply voltages to an assembly, use the ‘’Voltages and Signals’’ chapter to check the voltages at the assembly.

Cables can cause intermittent hardware failures.

Noise or spikes in the power supply can cause the analyzer to fail.

Measurements in this chapter are only approximate (usually 1 dB or 10%) unless stated otherwise.

Use chassis ground for all measurements in this chapter unless stated otherwise.

To determine your firmware version code, press [ System Utility ] [ MORE ] [ S/N VERSION ].

Logic levels in this chapter are either TTL level high or TTL level low unless stated otherwise. Toggling signal levels continually change from one TTL level to the other.

The troubleshooting tests in this chapter assume only one independent failure. Multiple failures can cause false results.

The troubleshooting procedures do not isolate failures to cables or connectors. If you suspect a cable or connector failure, check the device for continuity.

If you abort a self test before the self test is finished, the analyzer may fail its calibration routine. To prevent this from happening press [ Preset ] [ DO PRESET ] or cycle power after you abort the self test.

4-3