Adjusting the Analyzer

Agilent 35670A

To adjust common mode rejection

 

While monitoring the Y: value, adjust A2 R543 in the lower slot for a minimum marker value.

Disconnect the BNC cable from the analyzer’s CH 3 connector and connect to the CH 4 connector.

Press the the following keys:

[ Rtn ]

[ CHANNEL 4 ADJUSTMNT ] [ CMRR ]

Wait for the analyzer to set up the adjustment. The analyzer is ready when the adjustment message appears on the screen.

While monitoring the Y: value, adjust A2 R543 in the upper slot for a minimum marker value.

Set the power switch to off ( O ) and disconnect the minigrabber clips from A5 TP8.

5-16

Page 264
Image 264