Internal Test Descriptions

Agilent 35670A

Self-Test Descriptions

 

Self-Test Descriptions

Thirty-seven self tests are available that can be run in groups or individually. The following table lists the group of self tests that are run when you select

[ FUNCTIONL TESTS ], [ ALL ]. This group does not include any of the self tests that require a formatted flexible disk. The table lists the assemblies used by each self test and shows the assembly that would most likely cause the failure. To run these self tests in the order shown, press the following keys:

[ System Utility ]

[ SELF TEST ]

[ FUNCTIONL TESTS ] [ ALL ]

To run a single self test, press the softkey shown in the table instead of [ ALL ]. To determine the key path for the self-test softkeys, see “Self-Test Menu Map and GPIB Commands” starting on page 10-18.

Certain instrument malfunctions cause multiple self-test failures. Therefore, to determine the most likely cause when more than one self test fails, look in the ‘’Functional Tests All Self-Test Group’’ table for assemblies common to all failing self tests.

10-10

Page 392
Image 392
Agilent Technologies 35670-90066 manual Self-Test Descriptions