To run the program in semiautomated mode

Verifying Specifications

Agilent 35670A

To run the program in semiautomated mode

 

To run the program in semiautomated mode

You must have an GPIB printer connected to your system to run the program in semiautomated mode. If you do not have a printer, see ‘’To run the program without a printer’’ later in this chapter.

Press the following keys and when the program prompts you, type in the information for the title page of the test record and press [ ENTER ]:

[ TITLE PAGE ]

[ TEST FACILITY ]

[ FACILITY ADDRESS ] [ TESTED BY ]

[ REPORT NUMBER ] [ CUSTOMER ]

[ MORE ]

[ TEMP ]

[ HUMIDITY ]

[ LINE FREQUENCY ] [ RETURN ]

Press the following keys and when the program prompts you, type in the equipment configuration information:

[ EQUIP CONFIG ] [ AC CALIBRATO ] [ SYNTH. 1 ]

[ SYNTH. 2 ] (If needed)

[ LOW-D OSCILLATO ] (If needed)

[ MULTIMETER ] [ RETURN ]

The GPIB address is 100 x (interface select code) + (primary address). The interface select code for the test equipment and printer is 7 (for example, if the primary address is 8, the GPIB address is 708).

When entering the calibration due date, only four characters are displayed on the screen. However, you can enter up to nine characters and they will be printed.

3-8