Verifying Specifications

Agilent 35670A

To set up the source distortion test

 

To set up the source distortion test

Performance Test and Operation Verification

This test verifies that the Agilent 35670A meets its source output specification for harmonic and sub-harmonic distortion and spurious signals. In this test, the analyzer’s source is connected to its channel 1 input. The source is set for a maximum output level (5 Vpk) and the input range is set equal to the source level. The fundamental and harmonic is measured. The test records at the end of this chapter list the fundamental frequencies that are checked.

1 2 ch

1 4 ch

3-50

Page 108
Image 108
Agilent Technologies 35670-90066 manual To set up the source distortion test