Verifying Specifications

Agilent 35670A

Operation Verification Test Record - Two Channel

 

Serial Number:_______________________Report Number:____________________

Test Date:___/___/___

 

ICP Supply

 

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

 

 

 

 

 

 

 

 

 

 

Ch 1

Open Circuit Voltage

26 Vdc

32 Vdc

Vdc

 

 

 

 

 

 

 

Ch 2

Open Circuit Voltage

26 Vdc

32 Vdc

Vdc

 

 

 

 

 

 

 

Ch 1

Current

2.75 mA

5.75 mA

mA

 

 

 

 

 

 

 

Ch 2

Current

2.75 mA

5.75 mA

mA

 

Source Amplitude Accuracy

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

(%)

(%)

(%)

 

 

 

 

 

 

 

 

 

 

 

1 kHz, 0.1 Vpk

4

4

 

 

 

 

 

 

 

1 kHz, 3.0 Vpk

4

4

 

 

 

 

 

 

 

1 kHz, 5.0 Vpk

4

4

 

 

Source Flatness

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

(dB)

(dB)

(dB)

 

 

 

 

 

 

 

 

 

 

 

12.8 kHz

1

1

 

 

 

 

 

 

 

25.6 kHz

1

1

 

 

 

 

 

 

 

51.2 kHz

1

1

 

 

 

 

 

 

 

102.4 kHz

1

1

 

 

Source Distortion

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

(dBc)

(dBc)

 

 

 

 

 

 

 

 

 

 

 

12.8 kHz

 

60

 

 

 

 

 

 

 

51.2 kHz

 

40

 

 

 

 

 

 

 

102.4 kHz

 

40

 

 

10 of 10