Agilent Technologies 35670-90066 manual Source Amplitude Accuracy, Source Output Resistance, 19 of

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 149
Image 149
Ch 3 Current

 

Agilent 35670A

 

 

Verifying Specifications

 

 

 

Performance Test Record - Four Channel

Serial Number:_______________________Report Number:____________________

 

Test Date:___/___/___

 

 

 

 

 

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

 

 

 

 

 

 

 

 

 

 

 

Ch 2 Current

2.75 mA

5.75 mA

mA

 

 

 

 

 

 

 

 

Ch 3 Current

2.75 mA

5.75 mA

mA

 

 

 

 

 

 

 

 

Ch 4 Current

2.75 mA

5.75 mA

mA

 

Source Amplitude Accuracy

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

(%)

(%)

(%)

 

 

 

 

 

 

 

 

 

 

 

1 kHz, 0.1 Vpk

4

4

 

 

 

 

 

 

 

1 kHz, 3.0 Vpk

4

4

 

 

 

 

 

 

 

1 kHz, 5.0 Vpk

4

4

 

 

Source Output Resistance

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

(ohm)

(ohm)

 

 

 

 

 

 

 

 

 

 

 

Resistance

 

5

 

 

19 of 20

Page 149
Image 149
Agilent Technologies 35670-90066 Source Amplitude Accuracy, Source Output Resistance, Agilent 35670A, 19 of, Ch 2 Current