Agilent Technologies 35670-90066 How to troubleshoot the analyzer, Symptom, Troubleshooting Test

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 180
Image 180
How to troubleshoot the analyzer

Warning

Troubleshooting the Analyzer

Agilent 35670A

How to troubleshoot the analyzer

 

How to troubleshoot the analyzer

Review ‘’Safety Considerations’’ and ‘’Troubleshooting Hints.’’

Service must be performed by trained service personnel who are aware of the hazards involved (such as fire and electrical shock).

See chapter 6, ‘’Replacing Assemblies,’’ to determine how to disassemble and assemble the analyzer.

Determine initial test by comparing the analyzer’s symptoms to the symptoms in the following table.

Symptom

Troubleshooting Test

 

 

 

 

Screen blank †

Initial verification, page

Screen defective

4-5

After power up, >3 minutes before keys active

 

No response when key is pressed

 

Incorrect response when key is pressed

 

 

 

Keys are active and screen grid is displayed but screen is defective

Display, page 4-22

 

 

Error messages

Self tests, page 4-31

Calibration fails

 

Performance test fails

 

Intermittent failure

 

GPIB fails

 

Mic pwr fails

 

Serial port fails

 

Parallel port fails

 

External monitor does not work

 

 

 

External keyboard does not work

DIN connector, page

 

4-61

 

 

External trigger fails

Trigger, page 4-62

 

 

Nonvolatile states not saved after power cycled

Memory battery, page

Date display is ‘’Date: 01-01-BL’’

4-67

If the analyzer is failing, the grid may not appear for two minutes. Wait two minutes before assuming that a low level failure occurred.

Follow the recommended troubleshooting test until you locate the faulty assembly.

Replace the faulty assembly and follow the directions in ‘’What to do after replacing an assembly’’ in chapter 6, ‘’Replacing Assemblies.’’

4-4

Page 180
Image 180