Agilent Technologies 35670-90066 manual To set up the amplitude accuracy test, 1 2 ch, 1 4 ch

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 75
Image 75
To set up the amplitude accuracy test

Agilent 35670A

Verifying Specifications

 

To set up the amplitude accuracy test

To set up the amplitude accuracy test

Performance Test and Operation Verification

This test verifies that the Agilent 35670A meets its single channel amplitude specification for FFT full scale accuracy at 1 kHz. In this test, an ac calibrator outputs a 1 kHz signal with an exact amplitude to all channels. This test checks amplitude accuracy at 27, 19, 9, 1, −11, 27, 35, 43, and 51 dBVrms.

1 2 ch

1 4 ch

3-17

Page 75
Image 75
Agilent Technologies 35670-90066 manual To set up the amplitude accuracy test, 1 2 ch, 1 4 ch