Specifications

Agilent 35670A

Real Time Octave Analysis — Option 1D1

 

Real Time Octave Analysis — Option 1D1

Standards

Conforms to ANSI Standard S1.11 - 1986,

 

Order 3, Type 1-D, Extended and Optional

 

Frequency Ranges

 

Conforms to IEC 651-1979 Type 0 Impulse,

 

and ANSI S1.4

 

 

Frequency ranges (at centers)

 

Online (real time)

 

 

 

 

 

 

 

 

 

 

1 channel

2 channel

4 channel

 

1/1 octave

 

0.063 Hz to 16 kHz

0.063 Hz to 8 kHz

0.063 Hz to 4 kHz

 

1/3 octave

 

0.08 Hz to 40 kHz

0.08 Hz to 20 kHz

0.08 Hz to 10 kHz

 

1/12 octave

0.0997 Hz to 12.338 kHz

0.0997 Hz to 6.169 kHz

0.0997 Hz to 3.084 kHz

 

Capture playback

 

 

 

 

 

 

 

 

 

 

1 channel

2 channel

4 channel

 

1/1 octave

 

0.063 Hz to 16 kHz

0.063 Hz to 16 kHz

0.063 Hz to 16 kHz

 

1/3 octave

 

0.08 Hz to 31.5 kHz

0.08 Hz to 31.5 kHz

0.08 Hz to 31.5 kHz

 

1/12 octave

0.0997 Hz to 49.35 kHz

0.0997 Hz to 49.35 kHz

0.0997 Hz to 49.35 kHz

 

1 to 12 octaves can be measured and

 

 

 

displayed.

 

 

 

 

 

 

 

1/1, 1/3, and 1/12 octave true center frequencies related by the formula:

 

f (i+1)

= 2

1 n

; n = 1, 3 or 12;

 

 

 

f

(

i

)

 

 

 

 

 

 

 

 

 

 

 

 

 

Where 1000 Hz is the reference for 1/1, 1/3 octave, and 1000 x2(1/24)Hz is the reference for 1/12 octave. The marker returns the ANSI standard preferred frequencies.

Accuracy

±0.2 dB

1 second stable average

 

single tone at band center

 

 

 

Readings are taken from the linear total power spectrum bin. It is derived from sum of each filter.

1/3 octave dynamic range

>80 dB (typical) per ANSI S1.11 - 1986

2 second stable average, limited by input noise

 

level

 

1-16

Page 34
Image 34
Agilent Technologies 35670-90066 manual Real Time Octave Analysis Option 1D1, = 1, 3 or