Verifying Specifications

Agilent 35670A

To set up the input coupling test

 

To set up the input coupling test

Performance Test only

This test verifies that the Agilent 35670A meets its input specification for ac coupling rolloff. In this test, a frequency synthesizer outputs a 1 Hz signal to all channels. The signal is measured in both ac and dc coupled modes. The value measured in ac coupled mode is subtracted from the value measured in dc coupled mode to determine the ac coupling rolloff.

1 2 ch

1 4 ch

3-24

Page 82
Image 82
Agilent Technologies 35670-90066 manual To set up the input coupling test