Verifying Specifications

Features of the Program

The program can automatically create a printout similar to the test records at the back of this chapter.

The program can beep when equipment connections need to be changed.

The program can start the test sequence at any test in the operation verification or performance test list.

The program can stop after each measurement or alternatively, only if a failure occurs.

The program can be run in manual mode.

Test Duration

In semiautomated mode, the operation verification tests require approximately 11 2 hours and the performance tests require approximately 2 1 2 hours.

Calibration Cycle

To verify the Agilent 35670A Dynamic Signal Analyzer is meeting its published specifications, do the performance tests every 12 months.

Recommended Test Equipment

The equipment needed for operation verification and performance tests is listed on page 1-18.Other equipment may be substituted for the recommended model if it meets or exceeds the listed critical specifications.

Also, if you want the test record to be automatically printed, you need an GPIB printer. If you want the printer to automatically leave top and bottom margins on every page, enable perforation skip mode (see your printer’s manual for directions). If you do not have an GPIB printer you must record the results of each test in the test records. These test records may be reproduced without written permission of Agilent Technologies.