Troubleshooting the Analyzer

Agilent 35670A

To troubleshoot CPU, memory, and buses failures

 

To troubleshoot CPU, memory, and buses failures

Use this test to isolate the failure when the power-on LEDs show a fail code or the analyzer locks up during the power-up tests.

Step 1. Compare the LED fail code to the following table.

G

Set the power switch to off ( O ).

Set the power switch to on ( l ) while watching the power-on LEDs.

The power-on LEDs are on the A7 CPU assembly and are visible through the rear

panel. To see the LEDs easier, remove the seven screws holding the rear panel to

the analyzer and lean the rear panel back. This also gives you access to reset switch

SW2.

Determine the probable faulty assembly by comparing the power-on LEDs

fail code to the following table.

The power-on LEDs are showing a fail code when the LEDs display a code for

more than 4 seconds.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 