Troubleshooting the Analyzer

Agilent 35670A

To perform initial verification

 

Step 7. Check signals required for power up.

G

Using a logic probe, check the following signals.

 

 

 

 

Signal Name

Test Location

TTL State

Probable Faulty Assembly

 

 

 

 

 

 

 

 

 

 

 

PVALID

A7 P8 pin 3

High

A98 Power Supply

 

 

 

 

 

 

RSTn

A7 P8 pin 5

High

A7 CPU

 

 

 

 

 

 

CASn

A7 P7 pin 3

Toggling

A7 CPU

 

 

 

 

 

 

RASn

A7 P7 pin 4

Toggling

A7 CPU

 

 

 

 

 

 

VDATA

A7 P7 pin 7

Toggling

A7 CPU

 

 

 

 

 

 

ASn

A7 P6 pin 10

Toggling

A7 CPU

Using a logic probe, check that the following TTL signals are toggling just after power up.

The signals may stop toggling when the CPU assembly finishes the bootrom self tests.

Signal Name

Test Location

 

 

 

 

PASn

A7 P8 pin 7

 

 

PDSACK1n

A7 P8 pin 8

 

 

PDSACK0n

A7 P8 pin 9

 

 

PRW

A7 P8 pin 10

 

 

PDSn

A7 P8 pin 11

 

 

PA(26)

A7 P8 pin 12

 

 

PA(16)

A7 P8 pin 13

If the signals are not correct, the A7 CPU assembly is probably faulty.

Using a logic probe, check that A7 P7 pin 2 (SCL) and A7 P7 pin 1 (SDA) toggle TTL states at least twice just after power up.

If the signals are not correct, the A7 CPU assembly is probably faulty.

If the signals are correct, go to page 4-15,‘’To troubleshoot power-up failures.’’