To set up the flatness test

Verifying Specifications

Agilent 35670A

To set up the flatness test

 

To set up the flatness test

Performance Test and Operation Verification

This test verifies that the Agilent 35670A meets its single channel amplitude specification for FFT full scale flatness relative to 1 kHz. In this test, the ac calibrator outputs a signal with an exact amplitude to all channels. The test records at the end of this chapter list the amplitudes and frequencies that are checked.

1 2 ch

1 4 ch

3-18

Page 76
Image 76