Input Resistance

 

Agilent 35670A

 

 

Verifying Specifications

 

 

 

Performance Test Record - Two Channel

Serial Number:_______________________Report Number:____________________

 

Test Date:___/___/___

 

 

 

 

 

Input Resistance

 

 

 

 

 

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

(%)

(%)

(%)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 1

10

10

 

 

 

 

 

 

 

 

 

9 dBVrms, Ch 1

10

10

 

 

 

 

 

 

 

 

 

11 dBVrms, Ch 1

10

10

 

 

 

 

 

 

 

 

 

27 dBVrms, Ch 2

10

10

 

 

 

 

 

 

 

 

 

9 dBVrms, Ch 2

10

10

 

 

 

 

 

 

 

 

 

11 dBVrms, Ch 2

10

10

 

 

 

ICP Supply

 

 

 

 

 

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

 

 

 

 

 

 

 

 

 

 

 

Ch 1 Open Circuit Voltage

26 Vdc

32 Vdc

Vdc

 

 

 

 

 

 

 

 

Ch 2 Open Circuit Voltage

26 Vdc

32 Vdc

Vdc

 

 

 

 

 

 

 

 

Ch 1 Current

2.75 mA

5.75 mA

mA

 

 

 

 

 

 

 

 

Ch 2 Current

2.75 mA

5.75 mA

mA

 

 

Source Amplitude Accuracy

 

 

 

 

 

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

(%)

(%)

(%)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz, 0.1 Vpk

4

4

 

 

 

 

 

 

 

 

 

1 kHz, 3.0 Vpk

4

4

 

 

 

 

 

 

 

 

 

1 kHz, 5.0 Vpk

4

4

 

 

 

Source Output Resistance

 

 

 

 

 

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

 

(ohm)

(ohm)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Resistance

 

5

 

 

13 of 14