Troubleshooting the Analyzer

Agilent 35670A

To troubleshoot performance test failures

 

Failing Performance Test Troubleshooting Guide

 

Failing Performance Test

Probable Faulty Assembly

Adjustment

Troubleshooting Test

 

(in order of probabilty)

 

 

 

 

 

 

 

 

 

 

Spurious signals

A1/A2 Input

 

Input and ADC, page 4-51

one channel

A5

Analog

 

Four channel, page 4-54

 

 

 

 

 

Spurious signals

A2

Input

 

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

 

Spurious signals

A5

Analog

ADC gain, offset, and

Distortion, page 4-56

all channels

mechanical

reference, page 5-7

 

 

A6

Digital

 

 

 

 

 

 

Noise

A1/A2 Input

 

Input and ADC, page 4-51

one channel

A5

Analog

 

Four channel, page 4-54

 

 

 

 

 

Noise

A2

Input

 

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

 

Noise

A5

Analog

ADC gain, offset, and

Distortion, page 4-56

all channels

mechanical

reference, page 5-7

 

 

 

 

 

Cross talk

A1/A2 Input

 

Distortion, page 4-56

 

 

 

 

ICP supply

A1/A2 Input

 

 

 

 

 

 

 

Source dc offset

A5

Analog

 

 

 

 

 

 

 

Source amplitude accuracy

A5

Analog

 

 

 

 

 

 

 

Source flatness

A5

Analog

 

 

 

 

 

 

 

Source distortion

A5

Analog

 

 

 

 

 

 

 

Source output resistance

A5

Analog

 

 

4-44

Page 220
Image 220