Agilent Technologies 35670-90066 To run the program in manual mode

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 70
Image 70
To run the program in manual mode

Warning

Verifying Specifications

Agilent 35670A

To run the program in manual mode

 

To run the program in manual mode

Use this procedure if you want to run the program in manual mode. You will be prompted to set up all test equipment and you can check the analyzer’s setup state after each measurement.

Write in the information needed on the title page of the selected test record.

The test records are located near the back of this chapter and may be copied without written permission of Agilent Technologies.

Press the following keys and when the program prompts you, set all GPIB addresses to 0:

[ EQUIP CONFIG ] [ AC CALIBRATO ] [ SYNTH. 1 ]

[ SYNTH. 2 ] (If needed)

[ LOW-D OSCILLATO ] (If needed)

[ MULTIMETER ] [ RETURN ]

Press the following keys:

[ TEST CONFIG ]

[ PROCEDURE ]

[ OP_VERIFY ] or [ PERFORMAN ]

[ STOP AFTER ]

[ EACH MEASUREMENT ] [ RETURN ]

Press the following keys to start the test:

[ START TESTING ]

[ START BEGINNING ]

Now follow the directions on the display and record the measurement result in the selected test record after every measurement.

If you want to view the analyzer’s setup state, press [ BASIC ] [ Disp Format ] [ MEASURMNT STATE ] or [ INPUT STATE ]. To continue the program, press

[ BASIC ] [ DISPLAY SETUP ] [ LOWER ] [ RETURN ] [ CONTINUE ]. If you changed any instrument setup states, press [ RESTART TEST ] instead of [ CONTINUE ] to ensure accurate measurement results.

During the test, the program prompts you to change the test equipment connections. Always turn the ac calibrator output to OFF or STANDBY before changing test equipment connections. The ac calibrator can produce output voltages that could result in injury to personnel.

The directions on the display briefly tell you how to connect test equipment. For detailed illustrations of equipment setup, see the setup illustrations starting on the next page.

3-12

Page 70
Image 70
Agilent Technologies 35670-90066 To run the program in manual mode