Verifying Specifications

 

To load the program

To load the program

For information about the program’s softkeys, see the menu descriptions starting on page 3-51.

Set the Agilent 35670A Dynamic Signal Analyzer’s power switch to off ( O ), then connect the analyzer, test instruments, and printer using GPIB cables.

If you have the PC Style Keyboard, option 1CL, connect the keyboard to the analyzer using the keyboard cable (see ‘’To connect the optional keyboard’’ in chapter 2).

Insert the Semiautomated Performance Test Disk into the analyzer’s disk drive, then set the power switch to on ( l ).

After the analyzer finishes its power-up calibration routine, press the following keys:

[ Local/GPIB ]

[ SYSTEM CONTROLLR ] [ System Utility ]

[ MEMORY USAGE ]

[ REMOVE WATERFALL ] [ CONFIRM REMOVE ]

[ RETURN ] [ MORE ]

[ SERVICE TESTS ]

[ PERFRMANC TEST ]

Now go to one of the following procedures to continue:

‘’To run the program in semiautomated mode’’

‘’To run the program without a printer’’

‘’To run the program in manual mode’’