Adjusting the Analyzer

Agilent 35670A

To adjust filter flatness

 

For the standard two channel analyzer, do the following to adjust filter flatness:

GSet the power switch to on ( I ).

Press the following keys:

[ System Utility ] [ MORE ]

[ SERVICE TESTS ] [ ADJUSTMTS ]

[ CHANNEL 1 ADJUSTMNT ] [ 50 kHz ]

Wait for the analyzer to set up the adjustment. The analyzer is ready when the adjustment message appears on the screen.

While monitoring the Yr: value, adjust A1 R115 for a marker value of 0 0.1 dB.

Press the [ 100 kHz FLATNESS ] softkey.

Wait for the analyzer to set up the adjustment. The analyzer is ready when the adjustment message appears on the screen.

While monitoring the Yr: value, adjust A1 R235 for a marker value of 0 0.1 dB.

Press the following keys:

[ Rtn ]

[ CHANNEL 2 ADJUSTMNT ] [ 50 kHz FLATNESS ]

Wait for the analyzer to set up the adjustment. The analyzer is ready when the adjustment message appears on the screen.

While monitoring the Yr: value, adjust A1 R615 for a marker value of 0 0.1 dB.