Agilent 35670A

Verifying Specifications

 

To set up the tach function test

To set up the tach function test

Performance Test and Operation Verification

This test is only for Agilent 35670A’s with option 1D0, computed order tracking. This test verifies that the Agilent 35670A meets its tachometer specification for trigger level accuracy. In this test, a frequency synthesizer outputs a signal to the tachometer input and to channel 1. The analyzer makes an accurate order measurement on channel 1 to verify the trigger level and slope.

1 2 ch

Using an HP 3326A

1 4 ch

Using an HP 3326A

3-37

Page 95
Image 95
Agilent Technologies 35670-90066 manual To set up the tach function test