Agilent 35670A

Troubleshooting the Analyzer

 

To troubleshoot disk drive failures

To troubleshoot disk drive failures

G G

This test isolates disk drive failures to the A7 CPU, the A100 Disk Drive assembly, or the flexible disk.

Step 1. Check the disk controller on the A7 CPU assembly.

Press the following keys:

[ System Utility ]

[ MORE ]

[ SELF TEST ] [ TEST LOG ]

[ Rtn ]

[ FUNCTIONL TESTS ] [ I/O ]

[ INTERNAL DISK ]

[ DISK CONTROLLR ] [ DISK FIFO ]

If the disk controller test aborts and displays the message Mass Storage Unit not Present!!, the Disk Drive assembly, or the cable to the Disk Drive assembly is probably faulty. The A7 CPU assembly could also be faulty but it is less probable.

If the disk controller or disk FIFO test fails, the A7 CPU assembly is probably faulty.

Step 2. Check the Disk Drive assembly.

Insert a formatted flexible disk into the Disk Drive assembly and press the following keys:

[ RESTORE ]

[ RANDOM SEEK ] [ SEEK RECORD ]

(any number between 1 and 2771) [ READ ]

[ READ/WRITE ]

If any of the self tests failed, insert a new formatted disk and repeat the previous step.

If the disk drive self tests still fail, the A100 Disk Drive assembly is probably faulty.

If the self test aborts and displays the message Bad or unformatted media, the most likely cause of the failure is a bad flexible disk. The analyzer can use either LIF (Logical Interchange Format) or a DOS formatted flexible disk.

4-57

Page 233
Image 233
Agilent Technologies 35670-90066 manual To troubleshoot disk drive failures