Agilent Technologies 35670-90066 manual To connect the analyzer to an external monitor

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 51
Image 51

Agilent 35670A

Preparing the Analyzer for Use

 

To connect the analyzer to an external monitor

To connect the analyzer to an external monitor

The External Monitor connector is a 9-pin D female miniature connector that can interface with an external, multisync monitor. The monitor must be compatible with the 24.8 kHz line rate, 55 Hz frame rate, and TTL signals provided by the

Agilent 35670A. A SONY CPD-1302 monitor and a NEC Multisync 3D monitor with EZPIXpc† driver has been checked and found compatible with the Agilent 35670A external monitor mode operation.

Set the analyzer’s power switch to on ( l ).

Set the monitor’s power switch to on and configure the input and timing mode if necessary.

See the manual supplied with the monitor for information on configuring the monitor’s input and timing mode.

Connect the external monitor’s input cable to the analyzer’s rear panel EXT MONITOR connector.

A cable with a 9-pin connector option or an adapter to a 9-pin connector is required to connect the monitor to the Agilent 35670A.

Press the following keys to enable external mode: [ Disp Format ]

[ MORE ] [ MORE ]

[ EXT MON ON OFF ]

Pin Number

Signal Name

 

 

 

 

3R

4G

5B

Pin Number

Signal Name

 

 

 

 

8HSYNC

9VSYNC

1, 2, 6

GND

The EZPIXpc driver converts TTL video signals into RGB analog signals, drives 75 ohm coax cable, provides RGB composite sync or RGB sync on green, for monitors with RGB input capability. EZPIXpc, Covid, Inc., 1725 West 17th St, Tempe, Arizona 85281, 800-638-6104

2-13