Agilent Technologies 35670-90066 manual A11 Keyboard Controller, Signal Name

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 362
Image 362
A11 Keyboard Controller

Voltages and Signals

Agilent 35670A

A11 Keyboard Controller

 

A11 Keyboard Controller

The following table lists signals routed between the A11 Keyboard Controller assembly and the A7 CPU assembly. This table shows several things — if the assembly generates or uses the signal or voltage, and if a signal is bidirectional. A description of each signal follows the table.

Signal Name

Pin(s)

A7 P1

A11

 

 

 

 

BBEEPER

3

S

 

 

 

 

HRNGA

11

 

 

 

 

HRNGB

9

 

 

 

 

HRNGC

2

 

 

 

 

HRNGD

1

 

 

 

 

RESET

5

S

 

 

 

 

SCL

15

S

 

 

 

 

SDA

13

 

 

 

 

SINTFPn

7

S

 

 

 

 

+5 V

10

 

 

 

 

Gnd

6, 8, 12, 14, 16

 

 

 

 

Not Used

4

 

 

 

 

SThis assembly is the source of the signal.

• This assembly uses the signal.

This signal is bidirectional.

— This assembly does not use this signal.

BBEEPER

Buffered Beeper — This line controls the frequency of the front panel beeper tone.

HRNGA

Half Range A — In both the two channel and four channel analyzer, a high on this line turns on

 

the channel 1 half range LED. This line goes high when the A1 or A2 Input assembly detects

 

that the amplitude of the channel 1 input signal reached half the set range.

HRNGB

Half Range B — In a two channel analyzer, a high on this line turns on the channel 2 half range

 

LED. In a four channel analyzer, a high on this line turns on the channel 3 half range LED.

 

This line goes high when the A1 Input assembly detects that the amplitude of the channel 2

 

input signal reached half the set range or when the A2 Input assembly detects that the

 

amplitude of the channel 3 input signal reached half the set range.

HRNGC

Half Range C — In a four channel analyzer, a high on this line turns on the channel 2 half

 

range LED. This line goes high when the A2 Input assembly detects that the amplitude of the

 

channel 2 input signal reached half the set range. This line is only used in four channel

 

analyzers.

HRNGD

Half Range D — In a four channel analyzer, a high on this line turns on the channel 4 half

 

range LED. This line goes high when the A2 Input assembly detects that the amplitude of the

 

channel 4 input signal reached half the set range. This line is only used in four channel

 

analyzers.

RESET

System Reset — A high on this line resets the digital logic on the A11 Keyboard Controller

 

assembly. This line pulses high during power-up and power-down, and when the A7 CPU

 

assembly’s microprocessor executes the RESET instruction or is externally reset.

9-18