Agilent 35670A

Verifying Specifications

 

To set up the external trigger test

To set up the external trigger test

Performance Test only

This test verifies that the Agilent 35670A meets its trigger specification for external trigger level accuracy. In this test, a frequency synthesizer outputs a 1 kHz signal to the external trigger input and a 12.8 kHz signal to channel 1. The analyzer makes an accurate triggered measurement on channel 1 to verify the trigger level and slope.

1 2 ch

Using an HP 3326A

1 4 ch

Using an HP 3326A

3-35

Page 93
Image 93
Agilent Technologies 35670-90066 manual To set up the external trigger test