Agilent Technologies 35670-90066 Self-TestTroubleshooting Guide continued, Failing Self Test

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 211
Image 211
Manual background

 

Agilent 35670A

 

Troubleshooting the Analyzer

 

 

 

 

To perform self tests

 

Self-Test Troubleshooting Guide (continued)

 

Failing Self Test

Probable Faulty Assembly

Adjustment

Troubleshooting Test

 

 

 

 

 

 

 

 

Input A-Wt Filter

A1/A2 Input

 

 

one channel or channel 1 and

 

 

 

3 or channel 2 and 4

 

 

 

 

 

 

 

 

 

Input A-Wt Filter

A6

Digital

 

 

all channels

 

 

 

 

 

 

 

 

Input AAF/Bypass

A1/A2 Input

Filter flatness, page 5-17

 

one channel or

 

 

 

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

 

Input AAF/Bypass

A6

Digital

 

 

all channels

 

 

 

 

 

 

 

 

Input ICP Source

A1/A2 Input

 

 

one channel or

 

 

 

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

 

Tachometer

A6

Digital

 

Tachometer, page 4-24

 

A10 Rear Panel

 

 

 

 

 

 

 

Source Filter

A5

Analog

 

 

 

 

 

 

Source DC

A5 Analog

 

 

 

 

 

 

Quick Confidence

A1/A2 Input

Input dc offset, page 5-10

Performance test, page 4-42

 

A5

Analog

ADC gain, offset, and

 

 

A6

Digital

reference, page 5-7

 

 

 

 

Filter flatness, page 5-17

 

 

 

 

 

 

All self tests pass

 

 

Next Step

 

† This test fails if a device is connected to the front panel input connectors.

‡ This test fails if the Source output is not connected to the Tachometer input.

4-35

Page 211
Image 211
Agilent Technologies 35670-90066 Self-TestTroubleshooting Guide continued, Failing Self Test, Probable Faulty Assembly