Preparing the Analyzer for Use

Agilent 35670A

To change the fuses

 

To change the fuses

Both fuses are installed at the factory.

Unplug the power cord from the analyzer.

Press in and turn the appropriate fuse holder cap counter-clockwise (use a small screw driver for the ac fuse). Remove when the fuse cap is free from the housing.

Pull the fuse from the fuse holder cap.

To reinstall, select the proper fuse and place in the fuse holder cap.

DC Fuse

HP 2110-0920 30 A 32 V Normal Blow

AC Fuse

HP 2110-0342 8 A 250 V Normal Blow

Place the fuse holder cap in the housing. Press in and turn clockwise.

2-10