Agilent 35670A

Verifying Specifications

 

 

To run the program without a printer

 

Now follow the directions on the display and record every measurement result

 

in the selected test record.

 

 

 

Warning

During the test, the program prompts you to change the test equipment

 

connections. Always turn the ac calibrator output to OFF or STANDBY before

 

changing test equipment connections. The ac calibrator can produce output

 

voltages that could result in injury to personnel.

 

 

 

 

The directions on the display briefly tell you how to connect test equipment. For detailed illustrations of equipment setup, see the setup illustrations starting on page 3-13.

If you want to pause the program and return the Agilent 35670A Dynamic Signal

Analyzer to front panel control, press [ BASIC ]. To continue the program, press

[ BASIC ] [ DISPLAY SETUP ] [ LOWER ] [ RETURN ] [ CONTINUE ]. If you changed any instrument setup states, press [ RESTART TEST ] instead of [ CONTINUE ] to ensure accurate measurement results.