What to do after replacing an assembly

Replacing Assemblies

Agilent 35670A

What to do after replacing an assembly

 

What to do after replacing an assembly

Reinstall all assemblies and cables that were removed during troubleshooting.

Do the required adjustments listed in the following table.

Do the self test, page 4-31.

Do the required performance tests listed in the following table.

Assembly Replaced

Required Adjustment

Required Performance Test

 

 

 

A1 Input

ADC gain, offset and reference, page 5-7

DC offset

 

Input dc offset, page 5-10

Noise

 

Common mode rejection, page 5-13

Spurious signals

 

Filter flatness, page 5-17

Amplitude accuracy

 

 

Flatness

 

 

Amplitude linearity

 

 

A-weight filter

 

 

Channel match

 

 

ICP supply

 

 

 

A2 Input

ADC gain, offset and reference, page 5-7

DC offset

 

Input dc offset, page 5-10

Noise

 

Common mode rejection, page 5-13

Spurious signals

 

Filter flatness, page 5-17

Amplitude accuracy

 

 

Flatness

 

 

Amplitude linearity

 

 

A-weight filter

 

 

Channel match

 

 

ICP supply

 

 

 

A5 Analog

Source, page 5-6

DC offset

 

ADC, page 5-7

Noise

 

Input dc offset, page 5-10

Spurious signals

 

 

Amplitude accuracy

 

 

Flatness

 

 

Amplitude linearity

 

 

Channel match

 

 

Single channel phase

 

 

accuracy

 

 

External trigger

 

 

Source amplitude accuracy

 

 

Source flatness

 

 

Source distortion

 

 

 

A6 Digital

 

Source amplitude accuracy

 

 

Source dc offset

 

 

Source flatness

 

 

 

6-4