Agilent 35670A

Adjusting the Analyzer

 

To adjust the input dc offset

For the standard two channel analyzer, do the following to adjust input dc offset:

GSet the power switch to on ( I ).

Press the following keys:

[ System Utility ] [ MORE ]

[ SERVICE TESTS ] [ ADJUSTMTS ]

[ CHANNEL 1 ADJUSTMNT ] [ OFFSET ]

Wait for the analyzer to set up the adjustment. The analyzer is ready when the adjustment message appears on the screen.

Adjust A1 R39 for a Y: reading of 90 dBVrms or less.

Press the following keys:

[ Rtn ]

[ CHANNEL 2 ADJUSTMNT ] [ OFFSET ]

Wait for the analyzer to set up the adjustment. The analyzer is ready when the adjustment message appears on the screen.

Adjust A1 R539 for a Y: reading of 90 dBVrms or less.