Agilent Technologies Agilent 86120C SNRSTATe, Turns the signal- to- noise calculation on and off

Models: Agilent 86120C

1 294
Download 294 pages 52.87 Kb
Page 201
Image 201
SNR[:STATe]

Programming Commands

CALCulate3 Subsystem

 

SNR[:STATe]

 

Turns the signal- to- noise calculation on and off.

Syntax

:CALCulate3:SNR[:STATe]{? {ON OFF 1 0}}

Attribute

Preset State: off

Summary

*RST State: off

 

SCPI Compliance: instrument specific

 

 

 

 

 

Note

 

Only one STATe command can be turned on at any one time. Attempting to turn

 

more than one state on at a time results in a “–221 Settings Conflict” error. Refer

 

to “Measure delta, drift, and signal-to-noise” on page 3-14 for additional informa-

 

tion on selecting measurements.

 

 

4-73

Page 201
Image 201
Agilent Technologies Agilent 86120C SNRSTATe, Turns the signal- to- noise calculation on and off, Preset State off, Syntax