Agilent Technologies Agilent 86120C manual WLIMitSTATe, Syntax, Attribute, Summary, Description

Models: Agilent 86120C

1 294
Download 294 pages 52.87 Kb
Page 165
Image 165
WLIMit[:STATe]

Programming Commands

CALCulate2 Subsystem

 

WLIMit[:STATe]

 

Turns wavelength limiting on and off.

Syntax

:CALCulate2:WLIMit[:STATe]{? {ON OFF 1 0}}

Attribute

Non- sequential command

Summary

Preset State: on

 

*RST State: on

 

SCPI Compliance: instrument specific

Description

When this function is on, the Agilent 86120C has an input range from

 

the WLIMit STARt to the WLIMit STOP. When this function is off, the

 

instrument displays peaks over the full wavelength range. The graphics

 

display always shows the range between WLIMit:STARt and

 

WLIMit:STOP, regardless of the state of this command.

 

Whenever the Agilent 86120C receives this command, it reprocesses

 

the data and performs a new peak search.

 

 

 

 

 

 

 

Non-sequential command

 

 

Always use an *OPC? query or a *WAI command to ensure that this command has

 

 

the time to complete before sending any more commands to the instrument. Refer

 

 

to “Always force the Agilent 86120C to wait for non-sequential commands” on

 

 

page 3-12for more information.

 

 

 

 

4-37

Page 165
Image 165
Agilent Technologies Agilent 86120C manual WLIMitSTATe, Syntax, Attribute, Summary, Description, Non-sequential command