Examples :CONF:ARR:POW :FETC:ARR:POW? :READ:ARR:POW? :MEAS:ARR:POW?

Programming Commands

Measurement Instructions

Examples :CONF:ARR:POW :FETC:ARR:POW? :READ:ARR:POW? :MEAS:ARR:POW?

:CONF:SCAL:POW -10 dBm :FETC:SCAL:POW? MAX :READ:SCAL:POW? MIN :MEAS:SCAL:POW? DEF

Query Response The following line is an example of a returned string when :MEAS:SCAL:POW? MAX is sent:

-5.88346500E+000

If six laser lines are located and :MEAS:ARR:POW? is sent, the follow- ing string could be returned. Notice that the first returned number indicates the number of laser- line values returned in the query. The measurement units are in dBm.

6,-1.37444400E+001,-1.10996100E+001,-9.62396600E+000, -7.94024500E+000,-7.01303200E+000,-1.04536200E+001