Agilent Technologies Agilent 86120C manual 18. Preset Values, PRESet

Models: Agilent 86120C

1 294
Download 294 pages 52.87 Kb
Page 224
Image 224
Table 4-18. Preset Values

Programming Commands

STATus Subsystem

 

PRESet

 

 

Presets the enable registers and the PTRansition and NTRansition fil-

 

ters.

 

Syntax

:STATus:PRESet

 

Attribute

Preset State: none

 

Summary

*RST State: none

 

 

SCPI Compliance: standard

 

 

Command Only

 

Description

The PRESet command is defined by SCPI to affect the enable register.

 

If you want to clear all event registers and queues, use the *CLS com-

 

mand.

 

 

Table 4-18. Preset Values

 

 

 

 

 

Status Node

Preset Value

 

 

 

 

Operation enable register

0

 

Questionable enable register

0

 

PTRansition filters

32767

 

NTRansition filters

0

Example

 

 

OUTPUT 720;”:STATUS:PRESET”

 