Syntax*RST

Programming Commands

Common Commands

*RST

The *RST (reset) command returns the Agilent 86120C to a known condition.

Syntax*RST

Description For a listing of reset conditions, refer to the following table. This com- mand cannot be issued as a query. Since this command places the instrument in single measurement acquisition mode, any current data is marked as invalid and a measurement query such as :FETCh? results in error number –230, “Data corrupt or stale”. You must ini- tiate a new sweep with :INIT:IMM before you can use the :FETCh com- mand.

Table 4-15. Conditions Set by *RST Reset (1 of 2)

Item

Setting

 

 

Display mode

single wavelength

Start wavelength

1270 nm

Stop wavelength

1650 nm

Graphical display

off

Measurement acquisition

single

Wavelength calibration

vacuum

Elevation correction value

0 meters

Wavelength units

nm

Amplitude units

dBm

Power offset

0 dB

Peak threshold

10 dB

Peak excursion

15 dB

Measurement speed

normal

4-8