:READ:ARR:POW:WAV?

Programming Commands

Measurement Instructions

<resolution>

MAXimum

0.01 resolution (fast update)

Constants

MINimum

0.001 resolution (normal)

 

 

DEFault

Current resolution

Examples

:CONF:ARR:POW:WAV DEF, MAX

 

:FETC:ARR:POW:WAV? DEF, MIN

 

:READ:ARR:POW:WAV?

 

 

:MEAS:ARR:POW:WAV?

 

 

:CONF:SCAL:POW:WAV 1300NM, MAX

 

:FETC:SCAL:POW:WAV? 1300NM, MIN

 

:READ:SCAL:POW:WAV? 1300NM

 

:MEAS:SCAL:POW:WAV? 1300NM

Query Response

The following line is an example of a returned string when

 

:MEAS:SCAL:POW:WAV? MAX is sent:

 

+1.5529258E-006

 

If six laser lines are located and :MEAS:ARR:POW:WAV? is sent, the following string could be returned. Notice that the first returned num- ber indicates the number of laser- line values returned in the query.

6,+1.54488100E-006,+1.54648400E-006,+1.54809000E-006,+1.54969900E-006,+1.55131100E- 006,+1.55292600E-006