Agilent Technologies Agilent 86120C POINts?, PTHReshold, Non-sequential command, Syntax, Summary

Models: Agilent 86120C

1 294
Download 294 pages 52.87 Kb
Page 162
Image 162
POINts?

Programming Commands

CALCulate2 Subsystem

Non-sequential command

Always use an *OPC? query or a *WAI command to ensure that this command has the time to complete before sending any more commands to the instrument. Refer to “Always force the Agilent 86120C to wait for non-sequential commands” on page 3-12for more information.

 

POINts?

 

Queries the number of points in the data set.

Syntax

:CALCulate2:POINts?

Attribute

Preset State: unaffected

Summary

*RST State: unaffected

 

SCPI Compliance: instrument specific

 

Query Only

Description

This is the number of points that will be returned by the

 

CALC2:DATA? query.

Query Response

For example, if six laser lines are located:

 

+6

PTHReshold

Sets the peak threshold limit used by the instrument to determine valid laser line peaks.

Syntax

:CALCulate2:PTHReshold{? {<integer> MINimum MAXimum DEFault}}

 

<integer> represents logarithmic units in dB. Valid range is 0 to 40.

 

Constant

Value

 

MINimum

0 dB

 

MAXimum

40 dB