Table 2-1. Test Equipment Required

TYPE

REQUIRED CHARACTERISTICS

USE

RECOMMENDED MODEL

Oscilloscope

Sensitivity: 1 mV

P,T

Agilent 1740A

 

Bandwidth: 20MHz & 100MHz

 

 

 

Input: differential, 50Ω & 10MΩ

 

 

 

 

 

 

RMS Voltmeter

True rms, 10MHz bandwidth

P

Agilent 3400B

 

Sensitivity: 1 mV

 

 

 

Accuracy: 5%

 

 

 

 

 

 

Logic Pulser

4.5 to 5.5Vdc @ 35mA

T

Agilent 546A

 

 

 

 

Multimeter

Resolution: 100nV

P,A,T

Agilent 3458A

 

Accuracy: 0.0035%, 6½ digit

 

 

 

 

 

 

CC PARD Test

No saturation at 30Adc

P

Tektronix P6303

Current Probe

Bandwidth: 20Hz to 20MHz

 

Probe/AM503 Amp/

 

 

 

 

TM500 Power Module

Electronic Load*

Voltage range: 60Vdc

P,A

Agilent 6060B

 

Current range: 60Adc

 

 

 

Power range: 300 watts

 

 

 

Open and short switches

 

 

CC PARD Test & IP

Value: 0.25 ohms >250W

P,A

 

Cal Resistive Load*

 

Accuracy: 1%

 

 

 

Rheostat or Resistor Bank

 

 

 

 

 

 

 

Current-Monitoring

6023A

 

P,A

 

Resistors

Value:

30mV @ 30A (1mΩ )

 

 

 

 

Accuracy: 1%

 

 

 

 

TC: 30ppm/°C

 

 

 

Value:

30mV @ 30A (1mΩ)

 

 

 

 

Accuracy: 0.05% **

 

 

 

6028A

TC: 30ppm/°C (A,P)

 

 

 

100MΩ ± 0.04% @ 25W

 

Guidline 9230/15

 

Value

 

 

 

Accuracy: 1%

 

 

 

 

PC: 0.0004% 1W

 

 

 

 

 

 

 

Calibration and Test

Value:

100Ω, 5%, 1W

A,T

 

Resistors

 

1Ω, 5%, 1/2W

 

 

 

 

1KΩ, 5% 1/4W

 

 

 

 

5KΩ, 5% 1/4W (6023A)

 

 

 

 

2KΩ, 0.01% 1/4W

 

 

 

 

 

 

Terminating

Value: 50Ω ±5%, noninductive

P

 

Resistors (4)

 

 

 

 

 

 

 

 

Blocking

Value: 0.01∝F, 100Vdc

P

 

Capacitors (2)

 

 

 

 

Common-mode

 

 

P

Ferrox-Cube

Toroidal Core

 

 

 

500T600-3C8,

 

 

 

 

Agilent 9170-0061

10